Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Beckman HD130B: CAL ADJ /5500 DATE: 02-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 74 CONFIGURATION: Fluke 5500A ============================================================================= # # Source: # Beckman Series 300, Series 3000 and HD Series # Hand-Held Digital Multimeters Service Manual (3000-940-143) # # Compatibility: # 5500/CAL or MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- J F W 1.002 HEAD EQUIPMENT SETUP 1.003 DISP Remove four screws and carefully lift off back cover. 1.004 DISP Turn the UUT on. 1.004 DISP Ambient temperature: 23øC - 27øC. 1.004 DISP Relative humidity: less than 80%. 1.004 DISP Allow the UUT to stabilize for at least 30 minutes. 1.005 HEAD DC VOLTAGE 1.006 DISP Connect the 5500A NORMAL Hi & Lo to UUT V-Ohm & COM. 1.007 RNG 200mV 1.008 5500 190mV S 2W 1.009 DISP Adjust R5, located at the center of the large circuit 1.009 DISP board, for a UUT reading of +190.0mV or +190.1mV. 1.010 5500 * S 1.011 HEAD AC VOLTAGE 1.012 RNG 200mV AC 1.013 5500 190mV 100H SI S 2W 1.014 DISP Adjust R14 (AC Full Scale), located in the upper left- 1.014 DISP hand corner of the large circuit board, for a 1.014 DISP reading of 190.0mV ñ1 count. 1.015 5500 * S 1.016 DISP Temporarily reinstall back cover of UUT. 1.017 RNG 2V AC 1.018 5500 1.900V 20kH SI S 2W 1.019 MEMI Enter UUT reading in volts: 1.020 MATH MEM1 = MEM 1.021 DISP Remove UUT back cover again. 1.022 MEMI Enter UUT reading in volts: 1.023 MATH MEM = 1.9 + (MEM - MEM1) 1.024 DISP Adjust C19 (High Frequency), located on right side of 1.024 DISP the small circiut board, for a UUT reading of [MEM]V. 1.025 DISP Temporarily reinstall back cover of UUT. 1.026 MEMI Enter UUT reading in volts: 1.027 MATH MEM1 = MEM - 1.9 1.028 JMPZ 1.030 0.02U 1.029 JMP 1.020 1.030 5500 * S 1.031 DISP Remove all connections and reinstall four screws 1.031 DISP securing back cover. 1.032 END